Is High Temporal Resolution Achievable for Paediatric Cardiac Acquisitions during Several Heart Beats? Illustration with Cardiac Phase Contrast Cine-MRI - Université de Lorraine Accéder directement au contenu
Article Dans Une Revue PLoS ONE Année : 2015

Is High Temporal Resolution Achievable for Paediatric Cardiac Acquisitions during Several Heart Beats? Illustration with Cardiac Phase Contrast Cine-MRI

Résumé

Background During paediatric cardiac Cine-MRI, data acquired during cycles of different lengths must be combined. Most of the time, Feinstein’s model is used to project multiple cardiac cycles of variable lengths into a mean cycle. Objective To assess the effect of Feinstein projection on temporal resolution of Cine-MRI. Methods 1/The temporal errors during Feinstein’s projection were computed in 306 cardiac cycles fully characterized by tissue Doppler imaging with 6-phase analysis (from a population of 7 children and young adults). 2/The effects of these temporal errors on tissue velocities were assessed by simulating typical tissue phase mapping acquisitions and reconstructions. 3/Myocardial velocities curves, extracted from high-resolution phase-contrast cine images, were compared for the 6 volunteers with lowest and highest heart rate variability, within a population of 36 young adults. Results 1/The mean of temporal misalignments was 30 ms over the cardiac cycle but reached 60 ms during early diastole. 2/During phase contrast MRI simulation, early diastole velocity peaks were diminished by 6.1 cm/s leading to virtual disappearance of isovolumic relaxation peaks. 3/The smoothing and erasing of isovolumic relaxation peaks was confirmed on tissue phase mapping velocity curves, between subjects with low and high heart rate variability (p = 0.05). Conclusions Feinstein cardiac model creates temporal misalignments that impair high temporal resolution phase contrast cine imaging when beat-to-beat heart rate is changing.
Fichier principal
Vignette du fichier
journal.pone.0143744.PDF (1.6 Mo) Télécharger le fichier
Origine : Fichiers éditeurs autorisés sur une archive ouverte
Loading...

Dates et versions

hal-01330830 , version 1 (29-11-2016)

Identifiants

Citer

Laurent Bonnemains, Freddy Odille, Christophe Meyer, Gabriella Hossu, Jacques Felblinger, et al.. Is High Temporal Resolution Achievable for Paediatric Cardiac Acquisitions during Several Heart Beats? Illustration with Cardiac Phase Contrast Cine-MRI. PLoS ONE, 2015, 10 (11), pp.e0143744. ⟨10.1371/journal.pone.0143744⟩. ⟨hal-01330830⟩
59 Consultations
111 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More