Accurate calibration and optimized measurement: use of an achromatic compensator in rotating polarizer spectroscopic ellipsometry

Abstract : We describe an improvement of a Rotating Polarizer Ellipsometer in order to increase the sensitivity for the measurement of a transparent sample. We added a compensator, namely a quasi-achromatic retardation quarter wave plate on the 400-700nm spectral range. This new configuration is named Polarizer Rotating-Polarizer Compensator Spectroscopic Ellipsometer (PRPCSE). We present a method for calibration and the advantages for measurement of the transparent materials. This method is applied to CuCl nanocrystals embedded in a NaCl transparent matrix.