Off-resonance effects and selectivity profiles in pulsed nitrogen-14 nuclear quadrupole resonance

Abstract : In order to alleviate base-line distortions in nitrogen-14 NQR spectra originating from pulse breakthrough, low power radio-frequency (rf) pulses were applied. It is recalled that the required power is four times lower than that for an equivalent NMR experiment. This is easily explained by the fact that, in NMR, half the amplitude of the rf field is active. Moreover, the selectivity profile (i.e. the peak amplitude as a function of the difference between the carrier frequency and the resonance frequency) exhibits a shape which is, in most cases, more favorable in NQR than in NMR. An appropriate theory has been developed for explaining these experimental observations. It is concluded that low power NQR is perfectly feasible and should even be recommended for most applications, provided that the line-width of the NQR signal is not too large.
Type de document :
Article dans une revue
Solid State Nuclear Magnetic Resonance, Elsevier, 2012, 47-48, pp.39 - 46. 〈10.1016/j.ssnmr.2012.08.004〉
Liste complète des métadonnées

https://hal.univ-lorraine.fr/hal-01477862
Contributeur : Sebastien Leclerc <>
Soumis le : lundi 27 février 2017 - 17:04:08
Dernière modification le : vendredi 17 août 2018 - 13:30:02

Identifiants

Collections

Citation

L. Guendouz, A. Robert, A. Retournard, S. Leclerc, S. Aissani, et al.. Off-resonance effects and selectivity profiles in pulsed nitrogen-14 nuclear quadrupole resonance. Solid State Nuclear Magnetic Resonance, Elsevier, 2012, 47-48, pp.39 - 46. 〈10.1016/j.ssnmr.2012.08.004〉. 〈hal-01477862〉