Microscopic gel-liquid interfaces supported by hollow microneedle array for voltammetric drug detection

Abstract : This report describes a method for integration of a gel-liquid interface in hollow microneedles compatible with minimally invasive, electrochemical detection of drugs in vivo. The electrochemical sensor was characterised using cyclic voltammetry with tetraethyl ammonium. The experimental work demonstrated the detection of propranolol as a representative drug in physiological buffer with the microneedle system. A calibration curve for propranolol was built from measurements with differential pulse stripping voltammetry, indicating a sensitivity of 43 nA mu M-1, a limit of detection of 50 nM and a linear range between 50 and 200 nM.
Type de document :
Article dans une revue
Sensors and Actuators B: Chemical, Elsevier, 2014, 201, pp.572-578. 〈10.1016/j.snb.2014.04.080〉
Domaine :
Liste complète des métadonnées

https://hal.univ-lorraine.fr/hal-01505217
Contributeur : Lcpme Ul <>
Soumis le : vendredi 9 novembre 2018 - 20:48:43
Dernière modification le : lundi 19 novembre 2018 - 14:28:47

Fichier

43-postprint.pdf
Fichiers produits par l'(les) auteur(s)

Identifiants

Collections

Citation

Patricia Vazquez, Grégoire Herzog, Conor O'Mahony, Joseph O'Brien, James Scully, et al.. Microscopic gel-liquid interfaces supported by hollow microneedle array for voltammetric drug detection. Sensors and Actuators B: Chemical, Elsevier, 2014, 201, pp.572-578. 〈10.1016/j.snb.2014.04.080〉. 〈hal-01505217〉