Dimensioning IRGA gas sampling systems: laboratory and field experiments

Abstract

Both laboratory and field experiments were carried out in order to define suitable configuration ranges for the gas sampling systems (GSSs) of infrared gas analyzers (IRGAs) used in eddy covariance measurements. In the laboratory, an original dynamic calibration bench was developed in order to test the frequency attenuation and pressure drop generated by filters. In the field, three IRGAs of the same type equipped with different filters or different rain caps were installed and run and the real frequency response of the complete setup was tested. The main results are as follows. – Filters may have a strong impact on the pressure drop in the GSS and this impact increases with flow rate. – Conversely, no impact of the tested filters on cutoff frequency was found, GSSs with and without filters presenting similar cutoff frequencies. – The main limiting factor of cutoff frequency in the field was found to be the rain cap design. In addition, the impact of this design on pressure drop was also found to be noteworthy.
Fichier principal
Vignette du fichier
amt-9-1361-2016.pdf (538.26 Ko) Télécharger le fichier
Origin : Publisher files allowed on an open archive
Loading...

Dates and versions

hal-01557776 , version 1 (06-07-2017)

Licence

Attribution

Identifiers

Cite

Marc Aubinet, Lilian Joly, Denis Loustau, Anne De ligne, Henri Chopin, et al.. Dimensioning IRGA gas sampling systems: laboratory and field experiments. Atmospheric Measurement Techniques, 2016, 9 (3), pp.1361 - 1367. ⟨10.5194/amt-9-1361-2016⟩. ⟨hal-01557776⟩
148 View
133 Download

Altmetric

Share

Gmail Facebook Twitter LinkedIn More